Table 3 Auto Planning setting list

From: Auto- versus human-driven plan in mediastinal Hodgkin lymphoma radiation treatment

Auto-planning Settings
Max Itaration 60     
Engine Type Biological     
Tuning Balance 5%     
Dose Fall-Off Margin 1.8 cm     
Hot-Spot Maximum Goal 102%     
Use Cold-Spot ROIs YES     
Target Optimization Goals
PTV 30 Gy     
CTV 30 Gy     
OARs Optimization Goals
ROI Type Dose (Gy) Volume (%) Priority Compromise
Heart Mean dose 3   Medium Yes
Heart Max DVH 7.7 50 Medium Yes
Heart Max DVH 15 33 Medium Yes
Left Lung Mean dose 5   Medium Yes
Left Lung Max DVH 5 15 Medium Yes
Left Lung Max DVH 10 12 Medium Yes
Left Lung Max DVH 20 10 Medium Yes
Right Lung Mean dose 5   Medium Yes
Right Lung Max DVH 5 15 Medium Yes
Right Lung Max DVH 10 12 Medium Yes
Right Lung Max DVH 20 10 Medium Yes
Esophagus Max DVH 28 50 Medium Yes
Esophagus Max DVH 30 50 Medium Yes
Thyroid Mean dose 20   Medium Yes
Thyroid Max DVH 18 50 Medium Yes
Thyroid Max DVH 25 33 Medium Yes
Left Breast Mean dose 0.3   High No
Left Breast Max DVH 2 4 High No
Left Breast Max DVH 5 2 High No
Right Breast Mean dose 0.3   High No
Right Breast Max DVH 2 4 High No
Right Breast Max DVH 5 2 High No
Spinal Cord Max dose 25   High Yes
PRV Spinal Cord Max dose 26   High Yes
Ring PTV (+ 1.5 cm) Max dose 15   High Yes
Body-PTV-Ring PTV Max dose 10   High Yes
  1. Abbreviations: ROI Region of Interest, PTV planning Target Volume, CTV Clinical Target Volume, OARs Organs At Risk, DVH Dose Volume Histogram