Table 1 Camera setting Templates

From: Evaluation of daily patient positioning for radiotherapy with a commercial 3D surface-imaging system (Catalyst™)

Thorax
 Tolerance Settings
  lateral 5.0 mm
  longitudinal 5.0 mm
  vertical 5.0 mm
  rotation 5.0°
  roll 5.0°
  pitch 5.0°
 Surface Settings
  Tolerance 10 mm
 Image Surface Averaging
  Time 3 s
Abdomen
 Tolerance Settings
  lateral 5.0 mm
  longitudinal 5.0 mm
  vertical 5.0 mm
  rotation 5.0°
  roll 5.0°
  pitch 5.0°
 Surface Settings
  Tolerance 15 mm
 Image Surface Averaging
  Time 4 s
Pelvis
 Tolerance Settings
  lateral 8.0 mm
  longitudinal 8.0 mm
  vertical 8.0 mm
  rotation 5.0°
  roll 5.0°
  pitch 5.0°
 Surface Settings
  Tolerance 15 mm
 Image Surface Averaging
  Time 3 s
  1. Predefined templates for camera settings with tolerance values regarding the depiction of surface deviations and surface averaging time