Table 3 Descriptive statistics of the paired measurements

From: Implementation of an intraoperative electron radiotherapy in vivo dosimetry program

% Absorbed dose MOSFET Film
Minimum 78.0 72.0
1st quartile 90.2 92.4
Median 92.5 97.3
2nd quartile 97.6 103.0
Maximum 111.6 123.4
Mean 93.9 97.9
Standard deviation 6.8 9.7
  1. Thirty pairs of values for which the prescribed dose was 90 % and the maximum absorbed dose delivered was 100 %