Table 2 Descriptive statistics of the valid measurements: 40 measurements taken with MOSFETs and 42 measurements taken with films

From: Implementation of an intraoperative electron radiotherapy in vivo dosimetry program

% Absorbed dose MOSFET Film
Minimum 78.0 72.0
1st quartile 90.0 92.7
Median 92.8 98.1
2nd quartile 97.9 103.1
Maximum 111.6 123.4
Mean 93.8 97.9
Standard deviation 6.7 9.0