Table 7 Summary of dosimetric measurements with a 2D detector array

From: The impact of flattening-filter-free beam technology on 3D conformal RT

  All Breast Neurocranium Lung Spine metastasis Prostate
Modulated fields 96 12 12 18 24 30
γ<1 Mean +SD [%] 99.7 ± 0.9 99.9 ± 0.4 99.1 ± 0.7 100.0 ± 0.0 99.5 ± 2.2 100.0 ± 0.0
  1. The table shows the number of measured multistatic fields, mean and standard deviation for γ<1 with the gamma criteria 3mm DTA (distance-to-agree) and 3% dose difference for doses above 5% of the dose maximum.