Table 1 Deviation analysis - single film, lateral motion

From: Scanned carbon beam irradiation of moving films: comparison of measured and calculated response

     ROIirradiated ROItarget
Figure 2 A [mm] φ [°] T [s] Mean [%] SD Mean [%] SD
a) 0    -2.30 4.92 -1.52 2.06
b) 8 0 3 -2.98 16.92 -0.07 6.31
c) 8 0 4 -2.83 15.56 -0.98 9.16
d) 8 0 5 -2.24 18.53 0.73 9.34
e) 15 0 3 -1.85 10.78 0.66 6.19
f) 15 90 3 -0.89 11.53 0.73 9.50
g) 15 0 4 -3.50 12.06 -0.36 6.74
h) 15 90 4 -4.39 11.19 -1.23 5.17
i) 15 0 5 -5.40 14.99 -1.45 7.75
k) 15 180 5 -5.74 12.74 -2.50 7.61
l) 15 270 5 -6.46 16.09 -2.03 6.35
m) 20 0 5 -7.32 21.52 -2.92 14.49
n) 20 90 5 -5.03 20.64 -0.13 13.62
o) 20 0 7 -4.80 21.76 -0.34 15.08
p) 20 90 7 -9.84 68.98 -2.94 17.02
Mean     -4.52 19.52 -0.92 9.60
SD     2.39 14.74 1.30 3.86
Minimum     -9.84 10.78 -2.94 5.17
Maximum     -0.89 68.98 0.73 17.02
  1. Statistical analysis of relative deviations between calculated and experimental film response in two regions-of-interest (ROI). The film response distributions are shown in Figure 2 for the 14 different motion parameter combinations with amplitude A, period T, and phase φ at start of irradiation.